HILIC Solutions #9: Sample Solvent in HILIC

by Giorgia Greco1 and Thomas Letzel2,

1Thermo Fisher Scientific, Germany

2Analytical Research Group, Technische Universität München, Germany.

The sample solvent, also called dissolution solvent or injection solvent, is the solvent used to prepare the solution in which the compound to be analysed by HPLC is dissolved. The choice of the sample solvent is fundamental for a successful separation, since this determines the initial retention of the compound on the stationary phase. Sample solvents with stronger elution strength than the initial conditions of the mobile phase cause a less effective retention with undesirable consequences, such as peak distortion, peak broadening and earlier elution.

It is good practice in the development of a HPLC method to dissolve the sample in the initial mobile phase or in a miscible weaker solvent. This is even more important in HILIC separation as it is generally more sensitive to the solvent strength mismatch between the sample solvent and the mobile phase in comparison to reversed-phase chromatography.

In HILIC, aqueous sample solvents have a high elution strength, and this can impede the partitioning of the analytes into the stationary phase. This effects predominantly the earlier eluting highly polar analytes, for which substantial peak distortion and decreased retention can be observed. Therefore, the lowest possible amount of water should be used for the preparation of the sample solvent, ideally pure acetonitrile.

However, this can cause solubility issues with highly polar analytes. A good approach is to prepare stock solutions in higher amounts of water. Typically water/acetonitrile (50:50, v/v ) are suitable for most applications. Then, working solutions can be prepared by dilution in the mobile phase, which usually contains 80–95% (v/v ) acetonitrile or in pure acetonitrile. If the solubility is still problematic with a high content of acetonitrile, working solutions may be prepared in mixtures of acetonitrile/isopropanol (50:50, v/v ).

It is important to remember that for methods including a needle wash step, the wash-solvent should also contain at least 80–90% (v/v ) acetonitrile, since part of the wash-solvent can be injected with the sample.

An alternative solution that can overcome the negative effect of solvent strength mismatch between the sample solvent and the mobile phase is the use of the minimum sample volume.
In the forthcoming instalment, we will move our focus from the pure chromatography and we will discuss the combination of HILIC separation with MS detection.

Next Tip: HILIC with MS detection

Dr. Giorgia Greco is currently working as an Application Scientist with Thermo Fisher Scientific in Germany focusing on LC-MS applications for food and environmental analysis. She received a PhD in Chemistry and worked as Post Doc researcher at the Technische Universität München, Germany. During her research, she specialized in the fundamental of LC-MS and in the separation and analysis of metabolites from human and food matrices, as well as organic contaminants in waste water samples, by hyphenated HPLC/MS and HILIC/MS techniques.

Thomas Letzel , Associate Professor, is head of the Analytical Research Group at the Technische Universität München, Germany. He received his PhD in Chemistry with Aerosol Analysis, worked as Post-Doc performing pharmaceutical analysis, built up his research group in bioanalysis with Habilitation in 2009 and extended his analytical experience from then in food and water analysis. In all areas he developed novel analytical platforms based on LC–MS for the characterization of organic molecules in complex matrices. Thereby techniques (such as HILIC or RP-UHPLC) are applied for new analytical solutions often in direct flow-coupling with (bio)functional assays. He is the author of more than 50 publications and two books.