Alternative Protein Testing: Method Validation for Regulatory Compliance

Discover how to perform alternative protein testing using validated analytical methods designed to meet EFSA and FDA standards, even with complex and novel food matrices.
| 4 min read
Food safety team reviewing documentation and test results in a laboratory setting, representing collaborative validation in alternative protein testing

iStock

Register for free to listen to this article
Listen with Speechify
0:00
4:00

As novel proteins from plant-based, fermentation-derived, and cultured sources enter the market, regulatory bodies are intensifying scrutiny on how these are tested. Alternative protein testing now demands rigorous method validation to meet European Food Safety Authority (EFSA) and U.S. Food and Drug Administration (FDA) requirements for accuracy, specificity, and reproducibility.

Beyond EFSA and FDA, global organizations such as Codex Alimentarius, the Food and Agriculture Organization (FAO), and the World Health Organization (WHO) are contributing to evolving frameworks that influence novel food validation worldwide.

For a broader overview of analytical strategies used throughout the production lifecycle, see our comprehensive hub article, Alternative Protein Technology: Analytical Strategies and Workflows for Novel Foods.

Essential Validation Metrics for Alternative Protein Testing

Validation must meet EFSA and FDA standards and cover key performance metrics:

  • Accuracy and precision: Confirm the method’s ability to produce correct and reproducible results.
  • Selectivity and specificity: Ensure the method can reliably differentiate target proteins from other components in the sample matrix.
  • Limit of detection (LOD) and limit of quantification (LOQ): Define the smallest amounts that can be reliably detected and quantified.
  • Linearity and range: Demonstrate a consistent analytical response across the concentration range of interest.
  • Robustness and ruggedness: Evaluate the method’s reliability under variable conditions and between different laboratories or analysts.

These metrics form the foundation of credible data and regulatory-ready methods. 

To support regulatory submissions, validated methods must comply with international standards such as Association of Official Agricultural Chemists (AOAC) guidelines and International Organization of Standardization (ISO) accreditation, ensuring scientific rigor and reproducibility.

Analytical Techniques by Alternative Protein Testing Stage 

Each stage in the alternative protein production process benefits from specific analytical tools tailored to its unique requirements. 

The following section outlines commonly applied techniques that align with regulatory expectations and scientific best practices at various stages. Together they represent a toolbox from which analysts select based on the product, process, and quality objectives.

Raw Material Assessment

Thorough characterization of raw materials is critical before production begins. The following methods help verify protein content, detect contaminants, and assess ingredient integrity:

  • High-performance liquid chromatography (HPLC): Quantifies protein and confirms chemical purity.
  • Enzyme-linked immunosorbent assay (ELISA): Detects allergens and target proteins.
  • Fourier-transform infrared (FTIR) spectroscopy: Provides rapid compositional fingerprinting.
  • Polymerase chain reaction (PCR): Identifies GMOs or contaminants at the molecular level.

These methods set a solid foundation for downstream quality.

In-Process Monitoring

Maintaining consistent quality during production demands real-time tracking. These analytical tools detect biochemical and environmental changes as they occur:

  • Near-infrared (NIR) and FTIR spectroscopy: Monitor compositional shifts and chemical changes in real time.
  • pH and dissolved oxygen (O₂) sensors: Maintain optimal environmental conditions during fermentation or cell culture.
  • Enzymatic assays: Deliver fast, specific insights into metabolite levels and biochemical activity.

Together, these techniques support continuous quality control and early detection of deviations during production.

Final Product QA/QC

The final product must meet strict safety, nutritional, and labeling requirements. The following methods validate that it is ready for market release:

  • Liquid chromatography with tandem mass spectrometry (LC–MS/MS): Provides high-sensitivity quantification of proteins and trace contaminants.
  • ELISA: Confirms the presence or absence of allergens with high specificity.
  • Total protein assays: Quantify overall protein concentration to verify nutritional claims.
  • Microbial testing: Ensures the product is free from pathogenic contamination.

Together, these tools confirm that the final product is safe, nutritionally accurate, and compliant with regulatory standards.

Building a Regulatory-Compliant Alternative Protein Testing Validation Plan

A well-structured validation plan ensures analytical methods perform reliably and meet EFSA and FDA regulatory expectations. Key components include:

  • Regulatory understanding: Address agency expectations with reproducible methods and traceable documentation.
  • ISO 17025 alignment: Apply standardized practices to meet traceability and data quality benchmarks.
  • Cross-functional collaboration: Involve QA, regulatory, and analytical teams to align scientific rigor with compliance.
  • Strategic outsourcing: Engage contract research organizations (CROs) or qualified vendors for method development and validation support.
  • Matrix-matched calibrations and internal standards: Compensate for matrix variability and improve method accuracy.

By following these practices, laboratories strengthen their scientific credibility, streamline regulatory interactions, and reduce risks of delays in product development or approval.

Validation Workflow for Alternative Protein Testing

A structured workflow ensures consistency, reliability, and regulatory compliance throughout the validation process. The following stepwise approach supports robust method development for alternative protein testing:

  1. Preliminary method development: Screen and optimize analytical conditions.

  2. Pre-validation diagnostics: Identify and address performance gaps before formal validation.

  3. Formal validation: Execute standardized protocols under controlled conditions.

  4. Documentation: Maintain complete records—protocols, SOPs, validation reports, and change logs.

This stepwise process enables defensible, repeatable validation across even the most complex protein matrices.

Best Practices and Pitfall Avoidance in Alternative Protein Testing

To stay efficient and audit-ready amid growing regulatory scrutiny, laboratories should embed the following best practices into their workflows:

  • Standardize protocols and calibrate regularly: Maintain method reproducibility by enforcing SOPs and routinely calibrating instruments.
  • Account for matrix effects: Alternative protein matrices often contain emulsifiers, colorants, or crosslinking agents that interfere with analyte extraction and detection. Use matrix-matched calibrations and spiked recovery samples to strengthen robustness and accuracy.
  • Train staff in regulatory expectations: Ensure teams understand and apply ISO 17025, EFSA, and FDA guidelines across all validation activities.
  • Participate in interlaboratory studies: Benchmark methods, assess reproducibility, and promote standardization across labs.
  • Document everything: Maintain detailed logs, including traceability records, SOP changes, and validation histories to support inspections and audits.

Avoid common pitfalls such as insufficient replication, unrealistic acceptance thresholds, or failure to account for matrix effects. These missteps often lead to non-compliance and costly rework.

Final Thoughts: Why Alternative Protein Testing Must Be Regulatory-Ready

In the fast-moving field of alternative proteins, validated testing is more than a regulatory checkbox—it is the backbone of scientific credibility and commercial viability. By embedding compliance considerations into every stage of product development, from raw material analysis to final product release, laboratories can ensure their methods withstand regulatory scrutiny.

As regulatory agencies tighten expectations and novel protein matrices grow more complex, proactive validation is essential to accelerating approvals, reducing costly setbacks, and delivering safe, trusted food products to market.