Characterizing Antibody Oligonucleotide Conjugates With Complementary LC–MS Workflows

Native SEC–MS, peptide mapping, and oligonucleotide analysis reveal different aspects of conjugation and payload integrity in a model AOC.
Written byAimee Cichocki
Presented byRoxana Eggleston-Rangel and Haichuan Liu, Ph.D.
Antibody-oligonucleotide conjugate visualization

Google Flow 2026

Register for free to listen to this article
Listen with Speechify
0:00
3:00

An average oligonucleotide-to-antibody ratio can describe the loading of an antibody–oligonucleotide conjugate (AOC), but it leaves several questions open. Which conjugate populations contribute to that average? Where does conjugation occur? Does the oligonucleotide payload retain its intended structure?

A recent Phenomenex symposium explored those questions using a trastuzumab biosimilar conjugated to an RNA duplex. Roxana Eggleston-Rangel, M.S., Senior Application Scientist at Phenomenex, examined the intact conjugate and payload using native size exclusion chromatography–mass spectrometry (SEC–MS). Haichuan Liu, Ph.D., Staff Scientist, Technical Marketing Biopharma at SCIEX, extended the analysis through peptide mapping and oligonucleotide LC–MS/MS.

The model conjugate provided a technical evaluation system for connecting measurements at different molecular levels. Together, the workflows moved from the distribution of intact species to the structural details of the antibody and its RNA payload.

Looking Beyond the Average Loading

Oligonucleotide-to-antibody ratio (OAR) describes how many oligonucleotide payloads attach to an antibody. Eggleston-Rangel explains that each addition changes the conjugate’s molecular weight and physicochemical properties, with potential consequences for stability and biological behavior.

Working in analytical science?

Register for a FREE Separation Science account to subscribe to the Separation Science Newsletter.

Subscribe for free

“Too many oligos can affect stability and cellular uptake, while too few may reduce therapeutic effectiveness,” cautions Eggleston-Rangel.

Native SEC separates molecules according to hydrodynamic size. In this study, the team used 50 mM ammonium acetate at pH 6.8. More heavily conjugated species eluted earlier: Eggleston-Rangel showed populations with OAR values of four, three, two, and one, while mass measurements supported the assignments.

Eggleston-Rangel describes the benefit of the coupling: “By coupling native SEC with mass spectrometry, we can preserve the intact conjugates and directly determine the molecular masses.”

Liu subsequently calculated an average OAR of approximately 1.7 from the relative abundances of the detected species. The distribution supplied context that the average alone could not convey, such as amounts and pressure of each OAR, showing the different loading states present in the model sample.

Examining the RNA Duplex and Free Strands

The size-based separation also provided access to smaller oligonucleotide components. Eggleston-Rangel described an earlier peak corresponding to intact siRNA duplex and a later peak corresponding to excess antisense strand. Mass spectrometry supported the duplex assignment and revealed minor variants, including low-level oxidation and strand dissociation.

“We successfully resolved multiple species and identified the oligonucleotide components within the sample,” notes Eggleston-Rangel.

This combination linked chromatographic behavior to molecular identity. SEC distinguished species by size, while MS helped establish what those species represented. The analysis therefore addressed both the conjugation profile and the condition of the RNA components.

Locating Changes Within the Antibody

Intact analysis established the loading profile, but Liu turned to peptide mapping to investigate the antibody in greater detail. A C18 separation resolved peptides from the antibody digest before tandem mass spectrometry.

“We obtain close to one hundred percent sequence coverage of both the heavy chain and light chain in just a single injection,” reports Liu.

Liu compared collision-induced dissociation (CID) with electron-activated dissociation (EAD). In the conjugation-site example, EAD generated a more even distribution of peptide fragments and additional diagnostic information from the modified group, strengthening the site assignment.

Continue reading below…
Application NotesThermo_Fisher_Scientific_TN
LC-UV Analysis of Large RNA Using an Inert UHPLC System
Explore solutions to the challenges of oligonucleotide analysis with the Thermo Scientific Vanquish Amplify UHPLC Platform's inert flow path.
Read More

The fragmentation methods also supplied different information about glycosylation. In the example Liu presented, CID preferentially removed the glycan, while EAD generated peptide fragments that retained it. “Based on this information, we can pinpoint the location of the glycans in the sequence,” advises Liu.

Further examples addressed disulfide bond mapping and deamidation-related isomers. Chromatography separated isomeric peptide peaks, and diagnostic EAD fragments helped distinguish aspartic acid from isoaspartic acid forms. These results illustrated how separation and fragmentation contribute different evidence to a structural assignment.

Identifying Payload Variants

Liu then examined the siRNA used for conjugate synthesis through ion-pair reversed-phase LC–MS/MS. The workflow separated sense and antisense strands and their impurities, followed by intact mass measurements and fragmentation for sequence confirmation.

The analysis distinguished the linker-bearing sense strand from a hydrolyzed species. Liu reported mass errors within 2 ppm for the oligonucleotide species shown and full sequence coverage for the sense and antisense strands in the CID experiment. The workflow also identified truncated impurities, including an n−3 antisense species and an n−1 sense species.

Liu extends the approach to the truncated species: “We can also apply the same strategy to confirm the impurities detected in these samples.”

These payload measurements added detail beyond the native SEC–MS profile. A size-based separation could distinguish a duplex from free strands; the ion-pair LC–MS/MS analysis investigated the identity and sequence of individual strands and their variants.

Matching Each Workflow to a Structural Question

The study connected three levels of AOC characterization on one MS platform. Native SEC–MS described intact loading populations and smaller components. Peptide mapping investigated conjugation sites and antibody modifications. Oligonucleotide LC–MS/MS examined payload sequences and impurities.

Together, these measurements give analysts a clearer basis for distinguishing differences in payload loading from changes within the antibody or RNA. That context helps turn a complex AOC profile into specific questions for further investigation during development and stability studies.

Add Separation Science as a preferred source on Google

Add Separation Science as a preferred Google source to see more of our trusted coverage

Meet the Author(s):

  • Aimee Cichocki is the Editorial Director at Separation Science and Chromatography Forum. Aimee brings a broad range of experience in creating, editing, and formatting scientific content. With a degree in medicinal chemistry, a 10-year background in formulation chemistry, an MBA, and a diverse background in publishing, Aimee guides editorial initiatives at Separation Science and Chromatography Forum. Aimee is dedicated to ensuring the delivery of informative, reliable, and practical content to our audience of analytical scientists.

    View Full Profile

Speakers

Here are some related topics that may interest you:

Loading Next Article...
Loading Next Article...