Combining molecular images can reveal relationships that individual techniques leave unresolved. But when the measurements come from adjacent tissue sections, differences in cellular composition and morphology can complicate their interpretation.
Yasmin Shanneik, PhD., of Baylor College of Medicine addresses this challenge during a US HUPO webinar. She describes a workflow that performs desorption electrospray ionization mass spectrometry imaging, or DESI-MSI, and spatial transcriptomics on the same tissue section.
The approach connects lipid and metabolite distributions with gene-expression patterns. Its development depended on establishing whether the first analysis preserved the material needed for the second.
Establishing Compatibility Between Measurements
DESI-MSI measures molecular distributions directly from a tissue surface under ambient conditions. Spatial transcriptomics maps gene expression within the tissue. Together, they offer complementary information about molecular differences across a sample.
Performing both analyses on one section reduces uncertainty about whether the datasets represent corresponding structures. However, their standard workflows required adjustment.
Shanneik describes a sequence that places DESI imaging first, followed by hematoxylin and eosin staining and then Visium spatial transcriptomics. Before applying the combined workflow to human samples, the team investigated whether DESI analysis affected RNA integrity.
Using mouse brain sections, researchers compared RNA quality across conditions that included DESI imaging and exposure to room temperature without imaging. Shanneik reports that RNA integrity remained comparable under the conditions tested, supporting further evaluation in human tissue.
The published study subsequently demonstrated the combined workflow in breast and lung cancer sections, reporting preservation of RNA quality and spatial transcriptomics performance after DESI-MSI.
For laboratories developing sequential analyses, this compatibility testing is a critical step. Preserving tissue morphology alone does not establish that the molecular material required for a subsequent assay remains suitable.
Connecting Metabolic and Transcriptional Patterns
Shanneik focused her presentation on breast cancer tissue. The researchers compared transcriptomic results from a section processed through the combined workflow with those from an adjacent section analyzed using the transcriptomics protocol alone.
She described similar gene-expression clustering patterns, providing further support for compatibility in the samples examined.
The next task was to align the datasets. Working with Aspect Analytics, a spatial biology software platform, the team brought the DESI images, histological information, and transcriptomic measurements into a common format for spatial comparison.
This enabled analysis of relationships between molecular signals within corresponding regions. Mapping metabolites and lipids adds a chemical perspective to the gene-expression data, enabling researchers to investigate associations across molecular classes.
In the example Shanneik presented, the combined measurements revealed molecular heterogeneity within tumor regions that appeared similar by routine histological assessment. The team could then examine which lipid signals associated with particular gene-expression patterns.
These correlations provided leads for further investigation. They did not, by themselves, establish a causal relationship or validate a clinical prediction.
Designing an Integrated Analytical Workflow
The study illustrates how multimodal analysis depends on more than combining datasets after acquisition. Analysis order, molecular preservation, and image registration all influence whether the resulting comparisons are meaningful.
Those decisions begin with sample preparation, where handling choices determine what remains available for measurement. In a sequential workflow, each step must also preserve the requirements of the analyses that follow.
Shanneik’s presentation showed how testing those dependencies can expand the information obtained from one section. By establishing compatibility and aligning the measurements, the researchers could investigate metabolic and transcriptional differences within the same tissue structures—and generate more focused questions about tumor heterogeneity.