PFAS laboratories face a growing menu of analytical methods. Yet these methods do not offer interchangeable routes to the same result. They target different matrices, measure different fractions of fluorinated contamination, and support different regulatory programs.

The first decision has little to do with instrumentation. Laboratories must establish what they need to measure, in which matrix, and for what purpose.

Methods 533 and 537.1: Drinking Water Analysis

EPA Methods 533 and 537.1 both use solid-phase extraction (SPE) followed by liquid chromatography–tandem mass spectrometry (LC-MS/MS). Together, they cover 29 PFAS in drinking water.

Method 533 measures 25 PFAS. Its analyte list emphasizes short-chain compounds and includes perfluorinated acids, sulfonates, fluorotelomers, and ether carboxylic acids. The method uses anion-exchange SPE and isotope-dilution quantitation, which helps laboratories compensate for losses and matrix effects.

Method 537.1 measures 18 PFAS, including legacy compounds and replacements such as HFPO-DA. It uses SPE with LC-MS/MS but follows a different extraction and calibration design.

The methods overlap, but neither replaces the other across every application. Laboratories should compare the required analyte list with each method rather than selecting a method based on the total number of compounds.

Working in analytical science?

Register for a FREE Separation Science account to subscribe to the Separation Science Newsletter.

Subscribe for free

EPA approves Methods 533 and 537.1 for monitoring under the PFAS National Primary Drinking Water Regulation. Other techniques may generate useful drinking water data, but they do not automatically satisfy federal compliance requirements.

Method 1633A: Multiple Environmental Matrices

EPA Method 1633A expands targeted analysis beyond drinking water. It covers 40 PFAS in:

  • Aqueous samples, including wastewater, surface water, and groundwater
  • Soil, sediment, and biosolids
  • Landfill leachate
  • Tissue, including fish tissue

The method uses LC-MS/MS and isotopically labelled standards. Matrix-specific preparation procedures address the different extraction and cleanup demands presented by water, solids, and tissue.

This breadth makes Method 1633A a strong choice for laboratories that receive several environmental sample types. It also brings operational demands. Laboratories must control background contamination, demonstrate acceptable performance, manage matrix-specific procedures, and meet the method’s extensive quality-control requirements.

Method 1633A therefore represents more than an expanded analyte list. It creates a complete workflow for generating comparable targeted PFAS data across complex environmental matrices.

Method 8327: A Performance-Based SW-846 Option

EPA Method 8327 sits within EPA’s SW-846 compendium. It uses LC-MS/MS to determine 24 PFAS in prepared samples or extracts and was validated using external-standard calibration.

The method supports analysis of non-drinking-water samples, including groundwater, surface water, and wastewater. Associated sample-preparation procedures may extend its application to other matrices.

As with other SW-846 methods, laboratories must understand the project requirements before use. SW-846 provides a performance-based framework, and the applicable permit, regulation, contract, or quality-assurance plan may impose additional instructions.

Method 8327 may suit projects built around established SW-846 procedures. However, laboratories should not treat it as a direct substitute for Method 1633A. The two methods differ in analyte coverage, calibration strategy, matrix scope, and quality-control design.

Method 1621: Screening for Adsorbable Organic Fluorine

EPA Method 1621 answers a different question. Instead of identifying individual PFAS, it measures adsorbable organic fluorine (AOF) in aqueous matrices using granular activated carbon adsorption and combustion ion chromatography.

The result provides an aggregate measure of organofluorine retained under the method conditions. PFAS may contribute to that result, but so can other fluorinated substances, including some pesticides and pharmaceuticals.

Continue reading below…
Expert ForumsFilling up a glass with clean drinking water from kitchen faucet.
Optimizing PFAS Sample Preparation: Best Practices for Solvent Evaporation, Throughput, and Reproducibility
Learn how solvent evaporation impacts PFAS sample preparation, including its effects on contamination risk, sample recovery, and reproducibility.
Read More

Method 1621 cannot determine which compounds produced the measured fluorine. It also does not capture every organofluorine compound with equal efficiency. EPA therefore describes AOF as a method-defined parameter.

This makes Method 1621 a screening tool rather than a replacement for targeted LC-MS/MS. Used alongside a targeted method, it can reveal a gap between the fluorine explained by measured PFAS and the broader adsorbable organic fluorine signal. That gap may support further investigation, but it does not prove the presence or concentration of unidentified PFAS.

Selecting the Right PFAS Method

The sample matrix provides the clearest starting point:

Analytical Need

Method to Consider

Regulated drinking water analysis

533 or 537.1

Broad short-chain PFAS coverage in drinking water

533

Targeted analysis across water, solids, biosolids, and tissue

1633A

PFAS analysis within an SW-846 project

8327

Aggregate organofluorine screening in aqueous samples

1621

The reporting objective then narrows the choice. Targeted methods quantify named compounds. Method 1621 measures a broader fluorine parameter without identifying its components. A laboratory may need both approaches when the project asks how much known PFAS is present and whether targeted analysis leaves a wider fluorine signal unexplained.

Finally, the governing program controls method acceptance. Laboratories should review the current regulation, permit, contract, and accreditation requirements before developing or validating a regulatory PFAS workflow.

Method selection defines more than the analyte list. It shapes sampling, extraction, calibration, quality control, reporting, and the conclusions that users can draw from the results.

This article was produced under Separation Science's AI Editorial Guidelines.

Frequently Asked Questions (FAQs)

  • What are EPA Methods 533 and 537.1?

    EPA Methods 533 and 537.1 are analytical methods used for drinking water analysis that utilize solid-phase extraction and liquid chromatography-tandem mass spectrometry to measure different quantities and types of PFAS compounds.

  • How do Methods 533 and 537.1 differ?

    Method 533 measures 25 PFAS emphasizing short-chain compounds, while Method 537.1 measures 18 PFAS, including legacy and replacement compounds. Each method has distinct extraction and calibration designs.

  • What is Method 1633 used for?

    Method 1633 is designed for a broader range of environmental matrices and can analyze 40 PFAS in various samples including wastewater, soil, sediment, and tissue, making it suitable for laboratories handling multiple sample types.

  • When should one use Method 8327?

    Method 8327 is best suited for projects that align with established SW-846 procedures, as it provides performance-based analysis for non-drinking-water samples like groundwater and wastewater, but it should not replace Method 1633A.

  • What is the purpose of Method 1621?

    Method 1621 serves as a screening tool to measure adsorbable organic fluorine in aqueous samples, providing an aggregate measure that may include but does not specifically identify PFAS compounds.

Add Separation Science as a preferred source on Google

Add Separation Science as a preferred Google source to see more of our trusted coverage

Meet the Author(s):

  • Aimee Cichocki is the Editorial Director at Separation Science and Chromatography Forum. Aimee brings a broad range of experience in creating, editing, and formatting scientific content. With a degree in medicinal chemistry, a 10-year background in formulation chemistry, an MBA, and a diverse background in publishing, Aimee guides editorial initiatives at Separation Science and Chromatography Forum. Aimee is dedicated to ensuring the delivery of informative, reliable, and practical content to our audience of analytical scientists.

    View Full Profile

Here are some related topics that may interest you:

Loading Next Article...
Loading Next Article...