Extraction of Vitamin B7 (Biotin) from Serum

Extraction of Vitamin B7 (Biotin) from serum Using EVOLUTE EXPRESS ABN prior to LC-MS/MS analysis.
Updated
Brought to you byBiotage and Separation Science
Register for free to listen to this article
Listen with Speechify
0:00
1:00

Biotage has produced an application note for the high reproducible extraction recoveries of vitamin B7 from serum while minimizing co-extractable material in the form of proteins, lipids and phospholipids.

Biotage_May17_App1.pngIntroduction
Vitamin B7 (Biotin) is a water soluble vitamin necessary for cell growth. Serum is extracted using the EVOLUTE® EXPRESS ABN 96-well plate. The application note provides details of the sample preparation procedure, together with UHPLC and mass spectrometry conditions.

Results and Discussion
Good retention and chromatographic peak shape was obtained using the C18-PFP column. Stripped serum was spiked at various concentrations from 25–5000 pg/mL for recovery determination. High reproducible recoveries > 80% with corresponding RSDs < 10% were demonstrated.

Post extraction residual phospholipid levels were investigated to provide an indication of extract cleanliness. The most abundant phospholipids in human serum (previously selected from full scan, SIR and precursor ion scanning experiments) were assessed using MRM transitions monitoring the common 184 product ion.

Extract cleanliness was also investigated using post-column infusion (PCI) experiments. Mobile phase and blank serum extracts were injected onto the LC-MS/MS setup while simultaneously infusing Vitamin B7, to determine regions of suppression.

Conclusion
EVOLUTE EXPRESS SPE products are shown to dramatically improve flow characteristics, and enhance sample preparation productivity providing clean, robust, sample preparation. This method can be automated using Biotage® Extrahera.™