From Chromatogram to Report: How to Review PFAS Results

PFAS data review must confirm peak identity, quantitation and quality-control performance before results reach the final report.
Written byAimee Cichocki
Laboratory staff reviewing PFAS data on computer screen

A thorough PFAS data review ensures accuracy in concentration calculations under EPA guidelines, tracing results back to raw data.

iStock

Register for free to listen to this article
Listen with Speechify
0:00
3:00

LC-MS/MS software can calculate a PFAS concentration within seconds. That number does not become a defensible result until the laboratory reviews the chromatography, calibration, quality controls, and sample history.

Data review should trace each reported value from the chromatogram to the final unit. The reviewer must determine whether the peak represents the target analyte, whether the method quantified it within a controlled range and whether any failure limits its use.

Begin With the Sample Record

Review starts before the chromatogram. The reviewer should confirm that the laboratory used the requested method and that the sample met requirements for container, preservation, holding time, and preparation.

Dilutions, subsampling, re-extractions, and deviations can change the reporting limit or interpretation. These actions must remain connected to the final result.

The EPA Method 1633A uses several criteria to establish identity, including signal-to-noise ratio, retention time, and ion ratio where a reliable confirmation transition exists.

The reviewer should examine:

  • Retention time against the calibration standard
  • Coelution with the labelled analogue
  • Quantifier and qualifier transitions
  • Ion ratio, where applicable
  • Linear and branched isomer pattern
  • Peak shape and baseline
  • Potential interference
  • Integration boundaries

These checks should use the raw chromatogram rather than a concentration table alone.

Working in analytical science?

Register for a FREE Separation Science account to subscribe to the Separation Science Newsletter.

Subscribe for free

Some PFAS lack a dependable confirmation ion. In these cases, retention time, labelled-standard behavior, and clean background carry greater weight.

Review Integration Consistently

Automated integration can fail when peaks sit near the reporting limit, overlap with background, or show unusual shapes. Manual integration may correct the result, but it also introduces analyst judgement.

The laboratory should define when manual integration is permitted and how reviewers document it. Analysts should apply the same approach to samples, standards, and quality controls.

A reviewer should question integrations that exclude part of an isomer pattern, shift the baseline to create a detection, or treat similar peaks differently across the batch.

Check the Quantitative Range

Results must fall within the established calibration range. A response above the highest standard requires an appropriate dilution or another action permitted by the method. A signal below the laboratory’s reporting threshold may require qualification or reporting as a non-detect, depending on the method and project.

Dilution changes more than the calculated concentration. It raises the effective reporting limit and may alter labelled-standard response. The reviewer should verify the dilution factor, final units, and method requirements before accepting the result.

Connect Results With Quality Control

The reviewer should assess each result against the controls that support it. A method-blank detection may affect low-level sample results. A failed labelled-standard recovery may affect only the target compounds associated with that standard. A failed calibration verification can affect samples analyzed since the previous passing check.

Matrix spikes and duplicates provide added context but do not automatically invalidate every result when they fail. The reviewer must determine whether the failure reflects general method performance, a sample-specific matrix effect, or sample heterogeneity.

This analyte-and-sample-level review avoids rejecting unaffected data or releasing results unsupported by the controls.

Assign Qualifiers With a Defined Meaning

A qualifier should tell the data user what limitation occurred. It should not serve as a generic warning attached to any unusual result.

Continue reading below…
SymposiaScientist's gloved hand holiding water sample in test tube from outdoor water source
2026 PFAS and Emerging Environmental Contaminants
Strengthen your PFAS analysis with practical strategies to address complex environmental challenges.
Read More

Laboratories should define qualifiers in the report and apply them according to the method, project plan, or regulatory program. Common uses include identifying estimated concentrations, blank-associated results, reporting-limit exceedances, and results that failed an identification or recovery criterion.

A qualified value remains a result with a stated limitation. A rejected value lacks sufficient support for its intended use. The laboratory should not use qualification to rescue data when the governing method requires reanalysis or rejection.

Verify Units and Calculations

PFAS results may be reported in ng/L for aqueous samples and ng/g for solids or tissue. Method 1633A reports soil, sediment, and biosolids on a dry-weight basis, while tissue results use wet weight unless project requirements specify otherwise.

The reviewer should verify sample mass or volume, final extract volume, dilution factors, percent solids, and unit conversions. These checks prevent a sound instrumental result from becoming an incorrect reported concentration.

Write a Useful Case Narrative

The final report should include enough context for the client or regulator to understand departures from normal performance. The case narrative should identify affected samples and analytes, describe the issue, record corrective action, and explain any remaining qualification.

It should distinguish between a laboratory problem and a sample-specific limitation. A failed matrix spike, for example, may show that the sample affected recovery even though the clean control passed.

The goal of review is not to make every result appear acceptable. It is to communicate what the laboratory measured, how well the method supported it, and which limitations the data user must consider.

This article was produced under Separation Science's AI Editorial Guidelines.

Frequently Asked Questions (FAQs)

  • What is involved in PFAS data review?

    PFAS data review involves confirming the use of the requested method, examining the sample history, ensuring proper integration of chromatograms, and validating results against quality controls and standard calibration.

  • How does LC-MS/MS software calculate PFAS concentration?

    LC-MS/MS software can quickly calculate a PFAS concentration, but the result must be reviewed by a laboratory to become defensible, ensuring all chromatographic data, calibration, quality controls, and sample history are thoroughly assessed.

  • What are the EPA guidelines for reviewing PFAS data?

    The EPA guidelines recommend checking for unresolved technical problems, anomalies in raw data, proper method selection, transcription accuracy, and ensuring results fall within the calibrated range during PFAS data review.

  • Why is verifying peak identity important in PFAS analysis?

    Verifying peak identity is crucial because a response in an MRM channel does not guarantee the presence of the target PFAS. Criteria like retention time and signal-to-noise ratio help to establish whether the peak represents the target compound.

  • How do we handle results that fall outside the calibration range?

    Results that exceed the calibration range must be appropriately diluted or dealt with as specified by the method. It’s essential to verify the dilution factor and ensure that the new reporting complies with method requirements.

Add Separation Science as a preferred source on Google

Add Separation Science as a preferred Google source to see more of our trusted coverage

Meet the Author(s):

  • Aimee Cichocki is the Editorial Director at Separation Science and Chromatography Forum. Aimee brings a broad range of experience in creating, editing, and formatting scientific content. With a degree in medicinal chemistry, a 10-year background in formulation chemistry, an MBA, and a diverse background in publishing, Aimee guides editorial initiatives at Separation Science and Chromatography Forum. Aimee is dedicated to ensuring the delivery of informative, reliable, and practical content to our audience of analytical scientists.

    View Full Profile

Here are some related topics that may interest you:

Loading Next Article...
Loading Next Article...