GC Pressure Errors in Splitless Injection: A Guide to Purge Flow

Understanding Purge Flow to the Split Vent on GCs can help you fix pressure errors and prevent contamination.
Close-up of a gas chromatograph inlet.

iStock

Register for free to listen to this article
Listen with Speechify
0:00
2:00

The dreaded pressure error; it's a common stumbling block in the lab, capable of derailing an entire day of analysis. As Chromatography Forum user ChrisT discovered, a common point of confusion for analysts performing splitless injections is the "purge flow to split vent" setting. While the term "splitless" might suggest a complete absence of flow, setting the value to 0 mL/min can lead to system errors and poor data quality.

Splitless injection is a technique used in gas chromatography to introduce a large volume of a dilute sample onto the column. This is achieved by closing the split vent for a short duration, ensuring that a high percentage of your analytes are transferred to the column.

Understanding Purge Flow to the Split Vent

The purge flow to the split vent is a crucial setting that takes effect after the initial splitless period. Once the purge time has elapsed (typically less than a minute), the split vent opens, and a purge flow sweeps the residual solvent and any high-boiling compounds out of the inlet liner. This prevents contamination and the appearance of unwanted ghost peaks in your chromatograms.

As forum contributor 70 eV explains, the split flow setting is for the flow rate that will be used after the injection, with the purge time dictating the injection's duration. Another user, Peter Apps, clarifies that the instrument manages the splitless transfer by re-routing the gas flow automatically, meaning the instrument takes care of it once the "splitless injection" option is selected.

Key point: The instrument needs this flow to maintain proper inlet pressure. Setting the purge flow to 0 mL/min can prevent the GC from establishing the correct pressure equilibrium, causing an error.

Troubleshooting GC Pressure Errors

If you're observing a pressure error during your splitless injection, consider these troubleshooting steps:

  1. Verify your inlet configuration: As user AICMM notes in the thread, ensure your method is correctly set to splitless injection in the software. Incorrect settings can cause the instrument to misinterpret your parameters.

  2. Set a non-zero purge flow: A small, non-zero value for the purge flow (e.g., 10-20 mL/min) is often necessary to maintain proper inlet pressure and avoid overpressure conditions.

  3. Optimize your purge time: Use a flow calculator or a pre-existing method to determine the optimal purge time for your specific application. As user Jake advises, various GC vendors have their own pressure flow calculators, in which users can enter values for specific instruments and methods.

By correctly implementing these expert-recommended strategies, you can prevent common GC pressure errors and ensure your splitless injections are both robust and reliable. Proper management of purge flow is a fundamental step toward achieving consistent and high-quality chromatographic data.

Meet the Author(s):

  • Shiama Thiageswaran is an Assistant Editor at Separation Science. She brings experience in academic publishing and technical writing, and supports the development and editing of scientific content. At Separation Science, she contributes to editorial planning and helps ensure the delivery of clear, accurate, and relevant information for the analytical science community.