High-Sensitivity Analysis of Residual Solvents in Drug Products – USP <467> Residual Solvents Procedure A

by | Oct 19, 2017

Data obtained from Class 1 and Class 2 standard solutions, in accordance with Water Soluble Articles, Procedure A, in USP <467> Residual Solvents

Shimadzu has produced an application note presenting data obtained from Class 1 and Class 2 standard solutions, in accordance with Water-Soluble Articles, Procedure A, in USP <467> Residual Solvents.

Shimadzu_High-Sensitivity_Analysis_of_Residual_Solvents_in_drug_products.pngIntroduction
Headspace GC methods specified in the USP (U.S. Pharmacopeia), General Chapters <467> Residual Solvents, are commonly used for analysis of residual solvents. This application note presents data obtained using the Shimadzu HS-20 Headspace Sampler and Nexis GC-2030 Gas Chromatograph, from Class 1 and Class 2 standard solutions, in accordance with Water-Soluble Articles, Procedure A, in USP <467> Residual Solvents.

Figure3.pngResults
For Class 1 standard solution, the system easily fulfilled the sensitivity requirements including the low-responding carbon tetrachloride. On the other hand, chromatographic resolution of acetonitrile and methylene chloride that is required for Class 2 standard mixture was completely achieved at a resolution of 2.4. 

Related Content

Advertisment

Advertisment

Advertisement

Advertisement