HPLC Mobile Phase Preparation and Buffer Selection: pH Control, Ionic Strength, Solubility Limits, and Degassing

Master HPLC mobile phase preparation and buffer selection with our comprehensive laboratory guide to pH control, buffering capacity, organic salt solubility limits, and degassing protocols under harmonized USP <621> rules.
Written byShiama Thiageswaran
Researcher using HPLC system for mobile phase preparation

Google Gemini

Register for free to listen to this article
Listen with Speechify
0:00
12:00

In high-performance liquid chromatography (HPLC) and ultra-high performance liquid chromatography (UHPLC), the mobile phase is far more than a passive carrier fluid. It directly governs retention selectivity (alpha), peak symmetry (USP tailing factor T), analyte ionization state, column stationary phase longevity, and optical baseline stability.

Yet, despite its critical influence on chromatographic data quality, mobile phase preparation remains one of the single most frequent sources of unexpected laboratory deviations. Inaccuracies in aqueous pH measurement, incorrect solvent mixing sequences, inadequate buffer capacity, or overlooking salt solubility thresholds in organic modifiers lead directly to drifting retention times, split peaks, elevated baseline noise, and catastrophic pump seal or column inlet blockages.

Mastering HPLC mobile phase preparation and buffer selection requires a rigorous understanding of solution physical chemistry: selecting buffers matched to analyte pKa, determining optimal ionic strength, preventing non-ideal solvent contraction and salt precipitation, and implementing degassing protocols that eliminate pump cavitation and baseline spikes.

Working in analytical science?

Register for a FREE Separation Science account to subscribe to the Separation Science Newsletter.

Subscribe for free

Quick Reference: Core Buffer Selection and Preparation Rules at a Glance

For bench scientists and method developers, these foundational rules govern compliant and reproducible HPLC mobile phase preparation:

  • The Effective Buffering Window: Always operate within +/- 1.0 pH unit of the buffer's pKa. Outside this window, buffering capacity drops by more than 90%, leaving the separation vulnerable to environmental shifts.
  • The Analyte Ionization Rule (The 2-pH Rule): To guarantee a single, uniform ionization state (either 99% ionized or 99% unionized), maintain the mobile phase pH at least 2.0 pH units above or below the analyte's pKa. Avoid method development within +/- 1.5 pH units of the analyte's pKa to prevent peak splitting and retention drift.
  • The "Measure Aqueous First" pH Rule: Calibrate the pH electrode with standard aqueous buffers and measure/adjust the pH of the pure aqueous buffer phase before adding organic modifiers, in strict compliance with USP <621> and Ph. Eur. 2.2.46.
  • Buffer Concentration Limits: Maintain buffer salt concentration between 10 mM and 50 mM for reversed-phase columns. Below 10 mM risks capacity collapse; above 50 mM elevates backpressure and salt precipitation risks in high-organic gradients.
  • Independent Volumetric Measurement: Always measure aqueous and organic solvent portions in separate graduated glassware prior to mixing to compensate for non-ideal excess volume of mixing (solvent contraction).
  • Solubility Thresholds in Acetonitrile: Never exceed 15 mM to 20 mM potassium phosphate when acetonitrile concentrations exceed 70% to 75% (v/v) during gradient ramps.

Physical Chemistry of Buffer Selection: Matching pKa and pH Control

A buffer is defined as an aqueous solution containing a conjugate weak acid-base pair that resists changes in hydrogen ion concentration ([H+]) upon the addition of small amounts of acid, base, or sample diluent.

What is the Henderson-Hasselbalch Equation in HPLC?

The quantitative behavior of an HPLC buffer system is described by the Henderson-Hasselbalch equation:

pH = pKa + log([A-] / [HA])

Where:

  • pH = Negative logarithm of hydrogen ion activity in solution.
  • pKa = Negative logarithm of the acid dissociation constant of the weak acid.
  • [A-] = Molar concentration of the conjugate base (deprotonated species).
  • [HA] = Molar concentration of the weak acid (protonated species).

Buffer Capacity (Beta) and the +/- 1.0 pH Operating Window

Buffer capacity (Beta) measures the moles of strong acid or base required to change the pH of one liter of buffer solution by one pH unit:

Beta = 2.303 * C * (Ka * [H+]) / ((Ka + [H+])^2)

Where C is the total molar buffer concentration ([HA] + [A-]), and Ka is the acid dissociation constant.

The Practical Rule of Thumb: Buffer capacity reaches its absolute theoretical maximum when pH = pKa (where [A-] = [HA]). As pH deviates from the pKa, buffering resistance declines exponentially:

  • At pH = pKa +/- 1.0 unit: The buffer retains approximately 33% of its maximum capacity.
  • At pH = pKa +/- 1.5 units: The buffer retains less than 10% of its capacity.
  • At pH = pKa +/- 2.0 units: Buffering capacity drops below 1%; the solution functions essentially as an unbuffered salt solution subject to continuous environmental drift.

Comprehensive Guide to HPLC and LC-MS Buffers

Selecting the correct buffer system requires aligning the buffer pKa with your target operating pH while verifying optical transparency and mass spectrometer compatibility:

Buffer Chemical System

Useful Buffering Range (pH)

Primary pKa Values (at 25 deg C)

UV Cutoff (nm)

Mass Spectrometry Compatibility

Primary Applications & Cautions

Trifluoroacetic Acid (TFA)

1.5 to 2.5

0.23

210 nm

Compatible with cautions

Excellent ion-pairing agent for peptide mapping; causes ESI signal suppression in LC-MS

Phosphate (H3PO4 / NaH2PO4)

1.6 to 3.1

2.15 (pKa1)

< 200 nm

Strictly Incompatible

Gold standard for low-UV optical detection (< 220 nm); non-volatile salt that fouls MS sources

Formate (Formic Acid / Ammonium Formate)

2.8 to 4.8

3.75

210 nm

Fully Compatible

Standard volatile buffer system for reversed-phase LC-MS and LC-MS/MS bioanalysis

Acetate (Acetic Acid / Ammonium Acetate)

3.8 to 5.8

4.76

210 nm

Fully Compatible

Robust volatile buffer for peptide, metabolite, small-molecule, and environmental testing

Phosphate (NaH2PO4 / Na2HPO4)

6.2 to 8.2

7.20 (pKa2)

< 200 nm

Strictly Incompatible

Compendial standard for neutral-to-mildly alkaline pharmaceutical QC assay release methods

Bicarbonate (Ammonium Bicarbonate)

6.8 to 8.8 / 9.3 to 11.3

6.35 (pKa1), 10.33 (pKa2)

210 nm

Fully Compatible

Volatile alkaline buffer; susceptible to CO2 outgassing and pH drift over long analytical batches

Ammonia (Ammonium Hydroxide)

8.3 to 10.3

9.25

< 200 nm

Fully Compatible

Clean volatile basic modifier; requires column packed with high-pH stable hybrid silica

Phosphate (Na2HPO4 / Na3PO4)

11.3 to 13.3

12.38 (pKa3)

< 200 nm

Strictly Incompatible

Highly alkaline; strictly requires polymeric or bridged-ethyl hybrid (BEH) stationary phases

Analyte Retention as a Function of Ionization: The 2-pH Rule

In reversed-phase HPLC, an uncharged (neutral) organic molecule exhibits substantially greater hydrophobic retention on alkyl-bonded stationary phases (such as C18 or C8) than its charged, ionized counterpart:

Weak Acids (R-COOH):

  • At pH < pKa - 2: The acid is fully protonated and neutral (R-COOH). Hydrophobic retention on reversed-phase columns is maximized.
  • At pH > pKa + 2: The acid is deprotonated into an anion (R-COO-). Hydration increases dramatically, and retention times plunge.

Weak Bases (R-NH2):

  • At pH < pKa - 2: The base is fully protonated into a cation (R-NH3+). Reversed-phase retention is minimized, and basic analytes risk secondary silanol interactions (see our guide on HPLC Peak Shape Troubleshooting).
  • At pH > pKa + 2: The base is uncharged and neutral (R-NH2). Hydrophobic retention is maximized.

The Operational Hazard: Developing an analytical method within the analyte transition zone (pH = pKa +/- 1.0) forces the compound to exist in a dynamic 50:50 equilibrium between charged and uncharged species. Minor fluctuations in mobile phase temperature, local column heating, or sample injection diluent will alter this equilibrium, producing severe split peaks, asymmetric tailing, and run-to-run retention time drift (see Fixing HPLC Retention Time Drift).

Ionic Strength: Balancing Peak Shape and Instrument Reliability

Ionic strength (I) quantifies the total concentration of ionic charges present in the mobile phase solution:

I = 0.5 * Sum(c_i * z_i^2)

Continue reading below…
Application NotesMedical research scientist typing sophisticated coding on his desktop computer in a biological applied science research jaboratory
How Modern LC/MS Strategies Drive Efficiency, Reduce Costs, and Improve Lab Performance
Learn how to maximize return on investment with the Agilent InfinityLab Pro iQ Series
Read More

Where c_i is the molar concentration of ion i, and z_i is the ionic charge number of that ion.

Suppressing Secondary Silanol Interactions on Silica Supports

On silica-based columns, residual silanol groups (Si-OH) undergo partial deprotonation above pH 3.5 to form negatively charged silanoate sites (Si-O-). Positively charged, basic analytes (R-NH3+) interact ionically with these acidic sites, leading to severe peak tailing (USP tailing factor T > 1.8) and plate efficiency loss.

Increasing mobile phase ionic strength introduces counter-ions (such as K+, Na+, or NH4+) that shield these silanoate sites:

  • Below 5 mM Buffer Concentration: Counter-ion concentration is insufficient. Basic analytes exhibit severe tailing, poor theoretical plate counts (N), and retention shifts dependent on sample injection mass.
  • Between 10 mM and 25 mM: The optimal operating sweet spot for most small-molecule reversed-phase methods. Silanol interactions are effectively shielded without increasing solution viscosity.
  • Above 50 mM: Diminishing chromatographic returns. Operating above 50 mM increases dynamic mobile phase viscosity, elevates system backpressure, accelerates wear on pump piston seals (see HPLC System Pressure Diagnostics), and dramatically increases the risk of salt precipitation during gradient runs.

Organic Modifier Miscibility and Salt Solubility Limits

The most damaging physical error in mobile phase execution is buffer salt precipitation inside pump heads, autosampler injection loops, or column inlet frits. While inorganic salts dissolve easily in pure water, their solubility falls drastically when organic modifiers are added.

The Solvent Dielectric Mechanism of Precipitation

Organic solvents possess significantly lower dielectric constants (epsilon) than pure water:

  • Pure Water: epsilon approx 80
  • Methanol: epsilon approx 33
  • Acetonitrile: epsilon approx 37
  • Tetrahydrofuran (THF): epsilon approx 7.6

Lowering the dielectric constant diminishes the solvent's ability to isolate and stabilize separated ions. Solvated ion pairs collapse into crystalline lattices. Acetonitrile has a substantially lower solubilizing capacity for inorganic salts than methanol due to differences in hydrogen-bonding networks.

Critical Solubility Thresholds for Reversed-Phase Work

To prevent in-line precipitation when combining aqueous buffers with organic modifiers:

  • Phosphate Buffers in Acetonitrile: Potassium phosphate salts (KH2PO4 / K2HPO4) precipitate in acetonitrile at concentrations above 15 mM to 20 mM when the acetonitrile fraction exceeds 70% to 75% (v/v). Sodium phosphate is even less soluble; sodium phosphate can precipitate at concentrations as low as 10 mM in 80% acetonitrile.
  • Phosphate Buffers in Methanol: Methanol forms robust hydrogen bonds with dissolved ions. Sodium and potassium phosphates typically remain completely soluble up to 80% to 85% (v/v) methanol at 20 mM.
  • Volatile Buffers (Ammonium Formate / Ammonium Acetate): Exhibit vastly superior organic solubility compared to inorganic phosphates. Ammonium acetate typically remains soluble up to 90% to 95% acetonitrile at concentrations up to 20 mM.

The Bench Solubility Verification Test (The 4:1 Beaker Test)

Before mounting any new buffered mobile phase onto an automated gradient system, execute this 10-minute bench diagnostic:

  1. In a clean, transparent glass beaker or vial, combine 1 part of your aqueous buffer stock with 4 parts of your planned organic modifier (simulating an 80% organic gradient ceiling).

  2. Swirl gently and let the mixture sit undisturbed for 10 minutes at ambient temperature.

  3. Illuminate the vial against a dark background with a narrow light beam or laser pointer. If any cloudiness, haze, opalescence, or micro-precipitate appears (the Tyndall effect), the buffer concentration is too high for the intended organic gradient ceiling.

  4. Dilute the buffer salt concentration or lower the gradient high-organic wash step until the mixture remains completely optical-grade clear.

Operational Best Practices: Step-by-Step Preparation Protocols

Achieving run-to-run retention time reproducibility (percent relative standard deviation %RSD < 0.5%) requires rigorous adherence to standardized liquid-handling sequences.

A. The "Measure Aqueous First" pH Rule (Harmonized Compendial Mandate)

A pervasive source of inter-laboratory variability is measuring or adjusting mobile phase pH after adding the organic modifier:

  • The Physical Dilemma: Glass pH electrodes are calibrated against standard aqueous reference buffers where the activity coefficient of hydrogen ions and the liquid-junction potential are defined strictly in pure water. Adding organic solvents (acetonitrile or methanol) alters the solvent dielectric constant, shifts the pKa of both the buffer and the electrode reference junction, and destabilizes the glass membrane potential. This yields an "apparent pH" that cannot be reproduced across laboratories.
  • The Harmonized Rule: Under USP <621> and Ph. Eur. 2.2.46, the pH of a mobile phase is defined strictly as the pH of the aqueous component prior to the addition of organic modifiers. Always prepare, titrate, and verify the buffer pH in 100% water before combining with organic solvents.

B. Managing Non-Ideal Volumetric Contraction (Excess Volume of Mixing)

When water and organic solvents are combined, thermodynamic intermolecular hydrogen bonding causes the mixture to compress:

  • Mixing 500 mL of pure water with 500 mL of pure methanol does not yield 1,000 mL of solution; it yields approximately 965 mL due to volumetric contraction (a 3.5% volume loss).
  • Mixing water and acetonitrile yields approximately 980 mL per 1,000 mL nominal mixture.

Standard Laboratory Preparation Procedure:

  • Never pour solvent A into a graduated cylinder to a mark and top it up with solvent B.
  • Always measure the aqueous phase in one dedicated graduated cylinder and the organic phase in a separate graduated cylinder, then combine both into the mobile phase reservoir bottle.
  • Clearly document the method standard operating procedure (SOP) as volumetric ratios (e.g., "60:40 v/v water/acetonitrile measured independently").

C. Membrane Filtration: Protecting Pump Seals and Frits

All mobile phases—especially those containing dissolved buffer salts—must be vacuum-filtered prior to connection to the HPLC system:

  • Membrane Selection: Use 0.22 um or 0.45 um pore size membrane filters. Nylon membranes offer broad compatibility with aqueous/organic mixtures; regenerated cellulose (RC) or polytetrafluoroethylene (PTFE) are ideal for aggressive organic mixtures. Avoid cellulose acetate or nitrate membranes that shed polymeric fibers or break down in organic solvents.
  • UHPLC Requirements: For sub-2-micron UHPLC columns featuring 0.2 um to 0.5 um column frits, filtering all aqueous buffers through 0.22 um filters is mandatory to prevent premature inlet plugging and high backpressure spikes (see Preventing Extra-Column Band Broadening in HPLC and UHPLC Systems).

Mobile Phase Degassing: Physics, Technologies, and Optimization

Dissolved atmospheric gases (predominantly nitrogen and oxygen) in HPLC mobile phases represent a severe hydraulic and optical hazard.

The Physics of Cavitation, Bubble Spikes, and Baseline Noise

Under Henry's Law, the solubility of a gas in a liquid is directly proportional to the partial pressure of that gas above the liquid.

  • Pump Intake Cavitation: During the backward refill stroke of a high-pressure pump piston, the local pressure inside the pump chamber drops below atmospheric pressure. This localized vacuum pulls dissolved gases out of solution, forming microbubbles. The trapped bubble compresses during the forward stroke, preventing check valves from seating hermetically, which causes severe pressure ripple and flow rate delivery collapse.
  • Detector Optical Artifacts: As mobile phase exits the pressurized column bed, it experiences an abrupt pressure drop entering the detector flow cell. Dissolved gases outgas as microbubbles inside the optical path, scattering light and generating sharp, erratic vertical baseline spikes (see HPLC Baseline Troubleshooting).
  • UV Baseline Drift: Dissolved oxygen absorbs ultraviolet radiation below 250 nm. In gradient elution, oxygen solubility changes continuously as organic composition rises, producing a rolling, climbing baseline.

Degassing Technologies Compared

Degassing Technology

Primary Operating Mechanism

Key Advantages

Operational Limitations & Cautions

Inline Vacuum Degasser

Liquid flows through gas-permeable fluoropolymer tubing (e.g., Teflon AF) held under continuous vacuum (< 100 mbar)

Automated, fully continuous, highly reproducible, no ongoing helium costs; standard on modern HPLC

Requires annual preventive maintenance; vacuum chambers can weep or fail if internal membranes foul

Helium Sparging

Bubbling insoluble helium gas sweeps out dissolved N2 and O2 due to helium's near-zero liquid solubility

Most thorough deoxygenation; mandatory for reductive electrochemical detection

High ongoing gas cylinder cost; continuous sparging selectively evaporates volatile organic modifiers over time

Ultrasonic Bath Sonication

High-frequency acoustic cavitation agglomerates microbubbles so they rise to the surface

Inexpensive; effective for rapid post-filtration degassing of mobile phase bottles

Degassing is temporary; solvent completely re-saturates with atmospheric air within 15 to 30 minutes

Daily Aqueous Buffer Maintenance Protocols

Shelf Life of Aqueous Buffers: Unpreserved aqueous buffer solutions (especially neutral phosphate or acetate buffers at pH 5.0 to 7.5) promote bacterial and fungal growth within 24 to 48 hours. Microbial biofilms clog inlet frits and produce ghost peaks.

  • Standard Rule: Prepare uninhibited aqueous buffers fresh daily. Never "top off" an existing mobile phase bottle; discard remaining fluid, rinse thoroughly with HPLC-grade water and isopropanol, and refill with fresh buffer.
  • Storage Limits: If an aqueous buffer must be kept overnight, store at 2 to 8 deg C and allow it to equilibrate to room temperature before connecting to the instrument.

HPLC Mobile Phase Preparation Diagnostic Matrix

Observed Chromatographic Problem

Primary Root Cause

Diagnostic Verification Check

Targeted Corrective Action

Monotonic Early Retention Drift Across Batch

Mobile phase pH placed inside analyte buffer transition zone (pKa +/- 1.0)

A 0.1 unit change in mobile phase pH causes large shift in capacity factor (k')

Shift mobile phase pH to satisfy the 2-pH rule (>= 2.0 pH units away from analyte pKa)

Monotonic Late Retention Drift Across Batch

Volatile organic modifier evaporation caused by continuous helium sparging

Retention increases steadily across long runs (> 12 h) as organic modifier evaporates

Cap reservoir bottles; switch to inline vacuum degasser or brief 5-minute initial sparge

Sudden Pressure Spike During High-Organic Ramp

Inorganic buffer salt precipitation at column inlet

System backpressure spikes abruptly when gradient reaches >= 75% B; clears on pure water flush

Dilute buffer salt concentration to < 15 mM; switch from potassium phosphate to ammonium acetate

Severe Peak Tailing on Basic Compounds (T > 1.8)

Secondary silanol interactions from inadequate mobile phase ionic strength

Tailing occurs exclusively on basic analytes; neutral markers elute symmetrically

Increase buffer concentration from < 5 mM to 15 to 25 mM; lower pH < 3.0

Intermittent Vertical Spikes on Optical Baseline

Microbubbles outgassing in detector optical flow cell

Spikes disappear when modest backpressure (20 to 50 psi) is attached to detector waste

Ensure vacuum degasser is pulling < 100 mbar; install detector waste restrictor (UV only)

Ghost Peaks in Gradient Blank Runs

Microbial contamination in aged aqueous buffer

Ghost peak area increases directly with longer initial column re-equilibration time

Discard aged buffer; prepare fresh daily; clean glassware without detergent surfactants

Pump Pressure Ripple Exceeding 2%

Pump head cavitation caused by outgassing solvent

Rhythmic pressure oscillation mirroring piston cycle; bubbles in pump intake lines

Prime degasser and pump heads with degassed isopropanol; verify degasser vacuum pump

Run-to-Run Retention Variability Across Analysts

pH adjusted after adding organic modifier or incorrect volumetric mixing

Different analysts record variable retention times using identical nominal recipes

Enforce SOP: calibrate electrode in water, adjust pH in 100% aqueous, measure components separately

Frequently Asked Questions (FAQ): HPLC Mobile Phase Preparation and Buffer Selection

What happens if you measure HPLC mobile phase pH after adding organic modifier?

Measuring pH after adding organic modifiers (such as acetonitrile or methanol) introduces significant measurement error. Glass pH electrodes are calibrated against pure aqueous standards. Introducing organic solvents alters the solvent dielectric constant, shifts the apparent pKa of the weak acid, and disrupts the liquid-junction potential across the electrode membrane. This generates an apparent, non-reproducible pH reading. Under harmonized compendial rules (USP <621> and Ph. Eur. 2.2.46), mobile phase pH must always be measured and adjusted in 100% aqueous buffer prior to adding organic modifiers.

Continue reading below…
WebinarsSeparationScience_LC_Expert_210126
Expert Answers: LC Troubleshooting and Best Practices
Tap into expert LC guidance as Jim Grinias answers your questions on troubleshooting, method choices, and performance across workflows.
Read More

What is the difference between buffer capacity and buffer concentration?

Buffer concentration is the total molar quantity of the weak acid and its conjugate base in solution (for example, 20 mM total formate). Buffer capacity (Beta) is the quantitative ability of that solution to resist changes in pH when an acid or base is introduced. Buffer capacity depends on both total concentration and proximity to the buffer's pKa. A 50 mM buffer operating 2.0 pH units away from its pKa has negligible practical buffering capacity, whereas a 10 mM buffer operating exactly at its pKa provides robust pH stabilization.

What is the maximum organic concentration for potassium phosphate buffers?

When using potassium phosphate buffers (KH2PO4 / K2HPO4) in reversed-phase gradients with acetonitrile, keep the buffer concentration at or below 15 mM to 20 mM if the acetonitrile fraction exceeds 70% to 75% (v/v). Above 75% acetonitrile, inorganic phosphate salts rapidly precipitate out of solution, clogging column inlet frits and pump seals. If higher organic percentages are required for gradient washing, substitute phosphate with a volatile organic buffer like ammonium acetate, or replace acetonitrile with methanol.

How does volumetric contraction affect mobile phase accuracy?

When water and organic solvents are mixed, intermolecular hydrogen bonding causes the total solution volume to shrink (excess volume of mixing). Combining 500 mL of water and 500 mL of methanol produces only 965 mL of solution (a 3.5% volumetric contraction). If an analyst fills a 1,000 mL graduated cylinder with 500 mL of water and tops it off to the line with methanol, they will add roughly 535 mL of methanol, shifting the organic concentration by 3.5% and disrupting chromatographic retention times. Always measure solvent components in separate graduated glassware before combining.

Can ammonium bicarbonate be used for low-UV HPLC detection?

Ammonium bicarbonate has a UV absorbance cutoff around 210 nm to 215 nm, which can elevate background noise when monitoring compounds at low wavelengths (190 nm to 215 nm). Furthermore, ammonium bicarbonate solutions are susceptible to slow carbon dioxide (CO2) outgassing over time, which causes the mobile phase pH to drift upward during long analytical sequences. If operating below 215 nm, phosphoric acid/phosphate buffers are preferred for optical UV detection, while ammonium formate is preferred for volatile LC-MS applications.

Summary and Practical Next Steps

Reliable, reproducible chromatographic separations begin at the balance and the pH meter:

  1. Match Buffer pKa to Operating pH: Always select a buffer with a pKa within +/- 1.0 pH unit of your method target, and keep the mobile phase pH at least 2.0 pH units away from the analyte's pKa to guarantee a single ionization state.

  2. Standardize the pH Measurement Sequence: Always calibrate pH electrodes with aqueous reference standards and adjust buffer pH in 100% water before combining with organic modifiers.

  3. Respect Volumetric Non-Ideality: Measure aqueous and organic solvent volumes in separate graduated cylinders to avoid concentration errors caused by excess volume of mixing.

  4. Pre-Screen Organic Salt Solubility: Execute the 4:1 bench beaker test before running high-organic gradient methods to confirm that buffer salts remain completely soluble across the entire gradient profile.

  5. Ensure Continuous Degassing: Deploy inline vacuum degassers operating below 100 mbar to protect pump heads from cavitation, prevent check valve leakage, and eliminate optical baseline spikes.