This Edition of HPLC solutions is discussing the way to measure column efficiency for system suitability with gradients elution.
Updated
Written byJohn Dolan
Register for free to listen to this article
Listen with Speechify
0:00
2:00

Q: I have just developed a gradient HPLC method and need to set the acceptance criteria for system suitability. The column efficiency is one of the measurements we usually include in system suitability, but I don’t know which peak I should use to measure the efficiency. Can you help me?

JWD: Let’s review how we measure column efficiency (also called column plate number, N) under isocratic conditions. Most of us use one of the two formulas shown in the figure. We can use the baseline peak width, which is determined by measuring the distance at the baseline, wb, between tangents drawn to the sides of the peak. This is divided into the retention time, tR, squared, and multiplied by 16. Because drawing tangents is not convenient without printing the chromatogram, a common alternative is to measure the width at half the peak height, w0.5, and performing the same calculation with the exception of multiplying by 5.54 to account for the normal difference between the baseline and half-height width of a Gaussian peak. This also is the way most data systems calculate N. The plate number should be constant throughout the chromatogram, although early eluted peaks usually have somewhat smaller values due to extra-column effects. It really doesn’t matter which peak you choose for this measurement, but most workers choose to use one of the active ingredients of a pharmaceutical mixture or the target compound for a forensic, environmental, or chemical analysis method.

Working in analytical science?

Register for a FREE Separation Science account to subscribe to the Separation Science Newsletter.

Subscribe for free

Now let’s switch to gradients. With gradient elution, all the peaks in the chromatogram have approximately the same peak width, whereas with isocratic separation, the peak widths increase with retention time. What happens if we apply the isocratic plate number calculation to gradients? You can see that if retention increases while the peak width remains constant, the calculated plate number increases. This simply is not possible, so it tells us that we can’t calculate gradient column efficiency in this manner.

HPLC Solutions #116: Column Efficiency for System Suitability with Gradients
Figure 1

There isn’t an easy way to calculate N for gradients, so we have to take a different approach. Some workers use the isocratic formula for N with gradients, but always use it with the same peak. This still does not give a correct value of N, but at least it is consistent, so you should be able to observe when the plate number degrades. A better way is to use the peak width or width at half-height as a measure of column performance in gradient elution. You should still specify which peak to use for the measurement, but otherwise the peak width gives similar information about column quality for gradients as the plate number does for isocratic conditions. I would put a new column in the system and measure the peak width for the target peak under normal operating conditions. Then either watch the width as the column ages or put an aged column in and measure the peak width. Using these data, you should be able to put a limit on the peak width and use this as one of your system suitability criteria for gradient methods.

Continue Learning With John Dolan

Explore in-depth chromatography training courses from John Dolan on Analytical Training Solutions, designed to help practicing scientists strengthen method development, troubleshooting, and day-to-day decision-making in the lab.

View John Dolan’s chromatography courses >>

Add Separation Science as a preferred source on Google

Add Separation Science as a preferred Google source to see more of our trusted coverage

Meet the Author(s):

  • John Dolan

    John Dolan is considered to be one of the world’s top experts in HPLC. He has written more than 300 user-oriented articles on HPLC troubleshooting over the last 30 years, in addition to more than 100 peer-reviewed technical articles on HPLC and related techniques. His three books (co-authored with Lloyd Snyder), Troubleshooting HPLC Systems, Introduction to Modern Liquid Chromatography (3rd edn), and High-Performance Gradient Elution, are standard references on thousands of desks around the world. He has taught HPLC training classes around the world to more than 10,000 students.

    View Full Profile

Here are some related topics that may interest you:

Loading Next Article...
Loading Next Article...