Best Practices for HPLC Column Storage to Maximize Longevity

Proper HPLC column storage is essential to maintain performance and lifespan. Learn storage tips to avoid common pitfalls in chromatography.
Updated
Written byJohn Dolan
Scientists researching biochemical in a laboratory, Science experiment.

Chalirmpoj Pimpisarn

Register for free to listen to this article
Listen with Speechify
0:00
3:00

A reader asked me if she was in danger of shortening the lifetime of an HPLC column by removing it after each batch of samples and storing it. This also raised the related question about expectations of shorter column lifetimes if the same instrument was used for multiple methods, each with a different column.

Can Improper Storage Affect Column Longevity?

The simple answer is that most columns can be stored indefinitely without concern about deterioration. By “indefinitely,” I mean months, but probably not years – there are slow chemical changes that go on with any column over time. It really doesn’t matter whether you store the column on the instrument or in a box on the shelf – the environment in the column will be the same in either case, assuming the column is filled with the same solvent. There are a few general rules that should be followed, however.

Essential HPLC Column Storage Practices

Proper column storage practices are critical to maintain functionality and avoid premature deterioration. Following these basic storage guidelines can help you avoid common HPLC maintenance issues.

Working in analytical science?

Register for a FREE Separation Science account to subscribe to the Separation Science Newsletter.

Subscribe for free

Removing Salts and Buffers Before Storage

Leaving a column unattended while containing salts and buffers is not a good idea. If any mobile phase evaporation occurs, salts will be left behind, and these will be nearly impossible to redissolve. Also, it is not a good idea to store the HPLC system with buffers in it – they can be corrosive or leave solid deposits. So flush out the column with non-buffered mobile phase before you shut down the system. For example, if you are using 40/60 methanol/buffer, flush the system with 20-30 mL of 40/60 methanol/water to remove any buffer residues.

Daily Storage Considerations for Ion Pairing Reagents

My exception for the removal of buffers is when ion pairing is in use. Ion pair reagents are very slow to equilibrate – perhaps 20-50 or more column volumes (>50 mL) will be needed. For this reason, it may be inconvenient to re-equilibrate for each use when the column is used daily. Instead, leave the mobile phase in the column, remove the column and insert the plugs in the ends of the column, use a piece of jumper tubing to connect the autosampler and detector, and flush the HPLC hardware to remove any ion pair reagents or buffer. Column lifetimes are expected to be shorter anyway with ion pairing methods, so it isn’t likely that such storage practices will give noticeably shorter column lifetimes. An alternative is to leave the column in the system and set a very low flow rate (e.g., 0.1 mL/min) so that fresh mobile phase is flushed through the column continuously.

HPLC Solutions #79: Column Abuse 2: Column Storage


Using Organic Solvents for Long-Term Storage

Maintaining a column in organic solvent helps protect it from contaminants and ensures longevity. The column should be stored in at least 30% organic solvent. We’re assuming reversed-phase chromatography here, so that would generally be >30% methanol or acetonitrile. This is to keep microorganisms from growing in the column and system hardware. I find that it is most convenient to flush the column, as above, to remove buffers and salts, then convert to 100% organic, flush the column and leave the system for its next use.

Continue reading below…
Thought Leadershipsciex-280526-hero
Sterile but Toxic: Detecting Food Contaminants After Vacuum Refining
Quantify heat-stable cereulide by direct LC-MS/MS at 0.05 µg/kg across four lipid-rich food matrices.
Read More

When to Store Columns On or Off the System

Whether to keep a column on the system or store it off the system depends on factors like frequency of use and methods. It really doesn’t matter if the column is left attached to the HPLC system or not. Or whether it is removed and the system is used for other methods. The main practices are to remove buffers and salts, store in sufficient organic to prevent microbial growth, and avoid conditions that allow for mobile phase evaporation (see HPLC Solutions #78).

Continue Learning With John Dolan

Explore in-depth chromatography training courses from John Dolan on Analytical Training Solutions, designed to help practicing scientists strengthen method development, troubleshooting, and day-to-day decision-making in the lab.

View John Dolan’s chromatography courses >>

Add Separation Science as a preferred source on Google

Add Separation Science as a preferred Google source to see more of our trusted coverage

Meet the Author(s):

  • John Dolan

    John Dolan is considered to be one of the world’s top experts in HPLC. He has written more than 300 user-oriented articles on HPLC troubleshooting over the last 30 years, in addition to more than 100 peer-reviewed technical articles on HPLC and related techniques. His three books (co-authored with Lloyd Snyder), Troubleshooting HPLC Systems, Introduction to Modern Liquid Chromatography (3rd edn), and High-Performance Gradient Elution, are standard references on thousands of desks around the world. He has taught HPLC training classes around the world to more than 10,000 students.

    View Full Profile

Here are some related topics that may interest you:

Loading Next Article...
Loading Next Article...