Where’s My Pesticide?

This Edition of HPLC solutions is about analysis of pesticide in water.
Updated
Written byJohn Dolan
Register for free to listen to this article
Listen with Speechify
0:00
3:00

A reader is trying to figure out the fate of a pesticide in water. For the test, he takes a soil sample (e.g., 5 g) and adds 100 mL of an aqueous solution containing 10 ppm of a pesticide. He takes 1 mL aliquots of the liquid phase each hour and analyses for the pesticide until it can no longer be detected. At this point, he doesn’t know whether the pesticide is adsorbed onto the soil or something else has happened to it, such as degradation. He is also concerned that his data may not be valid because each time he removes a 1 mL sample, the total volume of the system is reduced. He asked for some advice on how to figure out what is going on.

First, I suspect that removal of 1 mL aliquots for sampling will have no impact on the result. This is a 1% reduction in volume and I expect that the magnitude of change he is looking for is much greater than this. However, you could mathematically correct for the removal each time it was done. For example, the first aliquot would represent 1/100 of the original solution, the second aliquot would be 1/99, the third 1/98, and so forth as the total solution volume was reduced by successive samplings. Another alternative would be to add a volume of water equal to the aliquot after each aliquot were taken. So remove 1 mL of sample and add back 1 mL of water and the total volume will stay constant.

Working in analytical science?

Register for a FREE Separation Science account to subscribe to the Separation Science Newsletter.

Subscribe for free

For sampling, I would suggest either filtering or centrifuging the sample to separate the liquid and solid portion, then the 1 mL sample could be removed. Then remix the solid and liquid and let it interact until the next sample cycle. This would avoid unintentional removal of soil with each aliquot and would simplify sample preparation prior to HPLC analysis.

To figure out where the pesticide ended up, the easiest thing to do would be to analyse the soil and see if it can be recovered from the soil. I’d filter the solution again, then take a sample of the soil (mud) and dry it to remove the water. You’ll want to be careful to use drying conditions that won’t degrade the pesticide, such as air drying at room temperature or vacuum drying. On the other hand, if the pesticide is known to be stable, oven-drying may be fine and will be much faster. Next, weigh an aliquot of the soil and extract it under conditions you know will (or suspect) will remove the pesticide from the soil, for example, with an organic solvent such as ethyl acetate. Analyse this solution and you can calculate how much pesticide was on the original 5 g of soil to calculate the mass balance.

Whenever I start on a project like this, I want to make the first run under conditions that will be easy to see what’s going on. For example, if you normally would expect the pesticide to be present at 10 ppm, but at this concentration, the peaks are quite small, increase the concentration to 100 ppm for the first cycle. And I wouldn’t worry about correcting for volume loss from the whole solution due to aliquotting. On the first pass you want to find out if your experimental strategy will work, what kind of problems will arise, and to get a general feel about what the results will be (e.g., ±10-20% probably is close enough for a first pass). Once you’ve figured out how to run the experiment and what results to expect, you can refine the process to get higher quality data. You certainly don’t want to spend time pre-analysing every nuance and working at trace levels until you know your experiments are likely to work.

Continue Learning With John Dolan

Explore in-depth chromatography training courses from John Dolan on Analytical Training Solutions, designed to help practicing scientists strengthen method development, troubleshooting, and day-to-day decision-making in the lab.

View John Dolan’s chromatography courses >>

Add Separation Science as a preferred source on Google

Add Separation Science as a preferred Google source to see more of our trusted coverage

Meet the Author(s):

  • John Dolan

    John Dolan is considered to be one of the world’s top experts in HPLC. He has written more than 300 user-oriented articles on HPLC troubleshooting over the last 30 years, in addition to more than 100 peer-reviewed technical articles on HPLC and related techniques. His three books (co-authored with Lloyd Snyder), Troubleshooting HPLC Systems, Introduction to Modern Liquid Chromatography (3rd edn), and High-Performance Gradient Elution, are standard references on thousands of desks around the world. He has taught HPLC training classes around the world to more than 10,000 students.

    View Full Profile

Here are some related topics that may interest you:

Loading Next Article...
Loading Next Article...