Method Adjustment vs Change Part 3: Column Dimensions

This Edition of HPLC is taking a closer look at some of the potential adjustments related to the column dimensions.
Updated
Written byJohn Dolan
Register for free to listen to this article
Listen with Speechify
0:00
1:00

In this article, we’ll take a closer look at some of the potential adjustments related to the column.

In the first two parts of this series, we took an overview of the concept (Part 1) then at specific guidelines for mobile-phase adjustment (Part 2). This article takes a closer look at some of the potential adjustments related to the column.

Column Length
The guidelines for adjustments to the column length are the same for the European Pharmacopoeia (EP) and the United States Pharmacopoeia (USP), as summarized in Table 1 – a change of ±70% in column length is allowed. For example, let’s consider the most popular column size, 150 x 4.6 mm i.d. This guideline would allow 70% x 150 = 105 mm of length adjustment. Thus, any column in the 45 to 255 mm range would be allowed. Of course, only certain column lengths are available, but these limits allow for a change to most of the common column lengths: 50, 75, 100, 150, and 250 mm.

To view the full article complete the form below:



Add Separation Science as a preferred source on Google

Add Separation Science as a preferred Google source to see more of our trusted coverage

Meet the Author(s):

  • John Dolan

    John Dolan is considered to be one of the world’s top experts in HPLC. He has written more than 300 user-oriented articles on HPLC troubleshooting over the last 30 years, in addition to more than 100 peer-reviewed technical articles on HPLC and related techniques. His three books (co-authored with Lloyd Snyder), Troubleshooting HPLC Systems, Introduction to Modern Liquid Chromatography (3rd edn), and High-Performance Gradient Elution, are standard references on thousands of desks around the world. He has taught HPLC training classes around the world to more than 10,000 students.

    View Full Profile

Here are some related topics that may interest you:

Loading Next Article...
Loading Next Article...