Navigating the Helium Shortage: Translating Gas Chromatography Methods to Hydrogen Carrier Gas

As global helium supplies become increasingly constrained, laboratories are utilizing thermodynamic method translation to successfully adapt established gas chromatography workflows to hydrogen carrier gas.
Updated
Written byShiama Thiageswaran
Laboratory gas chromatography setup featuring hydrogen gas cylinder

Google Gemini

Register for free to listen to this article
Listen with Speechify
0:00
2:00

For decades, helium has been the default carrier gas for modern analytical laboratories. It is inert, safe, and provides excellent separation efficiency. However, facing ongoing global supply shortages and volatile pricing, the analytical community is actively seeking sustainable alternatives. Hydrogen has emerged as the most viable replacement, but switching a heavily validated gas chromatography system from helium to hydrogen is not as simple as changing a cylinder. It requires rigorous thermodynamic method translation to ensure that the chemical resolution and retention times of established gas chromatography methods are preserved.

The Thermodynamics of Hydrogen in Gas Chromatography

The transition to hydrogen fundamentally changes the physical dynamics inside the separation column. According to the van Deemter equation, which relates a column's resolving power to the linear velocity of the carrier gas, hydrogen possesses a distinct thermodynamic advantage over helium in gas chromatography.

Hydrogen has a lower viscosity and a higher optimal linear velocity. More importantly, the van Deemter curve for hydrogen is much flatter at higher velocities. This means that gas chromatography systems utilizing hydrogen can be run at significantly faster flow rates without suffering the sharp drop in separation efficiency that occurs when pushing helium past its optimal speed. Consequently, laboratories that successfully translate their gas chromatography methods to hydrogen often experience the secondary benefit of reduced run times and higher sample throughput.

Working in analytical science?

Register for a FREE Separation Science account to subscribe to the Separation Science Newsletter.

Subscribe for free

Mathematical Modeling for Gas Chromatography Method Translation

Because hydrogen and helium interact differently with the stationary phase and the analytes under specific temperature and pressure conditions, simply applying a helium-based temperature gradient to a hydrogen-driven gas chromatography system will result in altered retention times and potentially swapped elution orders for co-eluting peaks.

To prevent this, analysts must use thermodynamic method translation. This process involves complex mathematical modeling to adapt the physical parameters of the gas chromatography method. By inputting the original helium flow rates, column dimensions, and temperature ramps into translation software, chemists can calculate the exact hydrogen head pressures and scaled temperature gradients required to replicate the original chromatography. Proper method translation ensures that the relative retention indices remain identical, meaning the analytes exit the gas chromatography column in the exact same order and pattern, just on a compressed timeline.

Hardware and Safety Considerations for Gas Chromatography

While the theoretical chemistry of method translation is robust, shifting a gas chromatography system to hydrogen requires acknowledging physical hardware and safety limitations. Hydrogen is highly flammable, prompting many laboratories to move away from high-pressure cylinders in favor of on-demand hydrogen gas generators, which only produce the gas as the gas chromatography instrument needs it.

Furthermore, when coupling gas chromatography to a mass spectrometer (GC-MS), the vacuum system must be evaluated. Hydrogen is harder to pump away than helium due to its smaller molecular size. Standard vacuum pumps on older gas chromatography systems may struggle to maintain the required high vacuum in the mass spectrometer source when utilizing hydrogen at optimal flow rates. Analysts must often adjust the column internal diameter, use a restrictor, or upgrade their pumping capacity to fully actualize the benefits of hydrogen carrier gas in modern gas chromatography workflows.

This article was produced under Separation Science's AI Editorial Policy.

Add Separation Science as a preferred source on Google

Add Separation Science as a preferred Google source to see more of our trusted coverage

Meet the Author(s):

Here are some related topics that may interest you:

Loading Next Article...
Loading Next Article...