The Flame Ionization Detector - Part 1

Find out more about the flame ionization detector, which is the most commonly used GC detector.
Updated
Written byMatthew Klee
Register for free to listen to this article
Listen with Speechify
0:00
1:00

In this technical article we initiate a discussion on the workhorse detector of gas chromatography: the flame ionization detector.

The flame ionization detector (FID) is the premier detector in gas chromatography. It has unique properties and performance that puts it above and beyond all other general-use detectors in gas chromatography (or any other form of chromatography, for that matter).
Topping the list of unique attributes are (1) unit carbon response and (2) wide linear operating range (up to 7 orders of magnitude). When combined with its other attributes of low cost, ease of use, speed of response, and ruggedness, it is no wonder why the FID is the premier detector of choice for GC. The FID is a mass sensitive detector.
That means that its response is proportional to the mass of carbon that passes through it. In that regard, FID response is stated in terms of picograms carbon per sec. Detection limits for FIDs are in the low pg C /sec.

By reading the full article you will learn more about the flame ionization detector, which is the most commonly used GC detector..

To view the full article complete the form below:



Add Separation Science as a preferred source on Google

Add Separation Science as a preferred Google source to see more of our trusted coverage

Meet the Author(s):

  • Matthew Klee

    Dr Matthew S. Klee is internationally recognized for contributions to the theory and practice of gas chromatography. His experience in chemical, pharmaceutical and instrument companies spans over 30 years. During this time, Dr Klee’s work has focused on elucidation and practical demonstration of the many processes involved with GC analysis, with the ultimate goal of improving the ease of use of GC systems, ruggedness of methods and overall quality of results. 

    View Full Profile

Here are some related topics that may interest you:

Loading Next Article...
Loading Next Article...