The Importance of Multiple-Charge Ion Precursors in Peptide MS/MS Sequencing

The importance of multiple-charge ion precursors to obtain best results in peptide MS/MS sequencing is discussed.
Updated
Written byO. David Sparkman
Register for free to listen to this article
Listen with Speechify
0:00
4:00

We typically think of a mass spectrum as consisting of a precursor ion at high m/z and a series of increasingly lower m/z product (fragment) ions. These types of mass spectra are shown in training courses and textbooks as didactic examples because they are easy to interpret “by hand”. A spectrum with these characteristics arises from the selection of a single-charge precursor ion. However, these teaching examples can be misleading to the beginning mass spectrometrist who may come to believe that any proper MS/MS experiment begins with selection of a single-charge precursor. In this edition of MS Solutions we will show that use of multiple-charge ion precursors is useful (and often essential), to obtain best results, particularly for peptide sequencing experiments.

There are three key user-settable parameters which control fragmentation efficiency in an electrospray LC/MS/MS method employing a tandem instrument:

  1. The collision energy of the precursor ion in the collision cell; this is controlled by the voltage drop across the cell; the greater the magnitude of the voltage drop, the higher the velocity of the precursor ions, and therefore the more energetic their collisions.
  2. The cross-sectional area of the collision gas; a larger cross-sectional area gas such as argon will result in greater fragmentation efficiency than, for example, helium. (Note: the user cannot control the collision gas type in quadrupole ion trap instruments.)
  3. The charge on the precursor ion; higher charge state precursors exhibit greater fragmentation efficiency. This is the focus of the remainder of this paper.

We typically draw a single-charge positive ion of a peptide with the proton residing on the N-terminus. However, this is inaccurate. We know that the proton must be able to move rather freely or we would not see production of the variety of structurally significant ions that we know occur in peptide MS/MS. We therefore refer to the proton on a single-charge peptide ion as a “mobile” proton and recognize that there is some probability of the proton residing on any nitrogen in the polypeptide chain: the N-terminus, any of the peptide-bond nitrogens, or nitrogens in the basic side-chains of arginine, lysine, and histidine. Figure 1 shows the mobile proton concept schematically. This peptide, when used as an MS/MS precursor ion, is likely to form a variety of structurally useful product ions.

MS Solutions #17: The Importance of Multiple-Charge Ion Precursors in Peptide MS/MS Sequencing
Figure 1: Schematic of the mobile proton concept.

Fragmentation in low-energy CAD is a result of charge-site driven mechanisms. Therefore, the adducted proton must exhibit mobility in order for single-charge peptide precursors to generate product ions representing various backbone bond cleavages, e.g. b 1, b 2, b 3, etc. An adduct proton located on the peptide backbone also explains the production of complementary ions, e.g. both y and b ions from the same bond cleavage. An example mechanism for formation of complementary product ions is shown in Figure 2. The charge site on this pentapeptide is on the peptide bond between residues 3 and 4 and cleavage of this bond results in production of both the y 2 and b 3 ions.

MS Solutions #17: The Importance of Multiple-Charge Ion Precursors in Peptide MS/MS Sequencing-2
Figure 2: Charge-site driven fragmentation of a pentapeptide showing importance of the mobile proton concept in production of complementary ions from the same bond cleavage. (Courtesy of Jack Throck Watson).

The most commonly used proteolytic enzyme in protein sequencing experiments is trypsin. Trypsin cleaves the peptide bond on the C-terminal side of lysine and arginine residues. This means that nearly all tryptic fragments exhibit either an arginine (R) or a lysine (K) on the C-terminus. In the case of a C-terminal arginine, the mobility of the proton is severely limited by the presence of the strongly basic side-chain. The proton of a single-charge ion is, in fact, believed to be sequestered on the arginine sidechain [Figure 3(a)]. This places the charge site remote from the peptide backbone, i.e., remote from the desired cleavage sites. As a result, single-charge precursors with a C-terminal arginine exhibit severely limited production of structurally useful product ions.

Working in analytical science?

Register for a FREE Separation Science account to subscribe to the Separation Science Newsletter.

Subscribe for free

[Figure 3(b)] illustrates that the solution to this problem is the selection of multiple-charge precursor ions. In this situation, while one proton is sequestered on the R side-chain, the other acts as a mobile proton and provides the necessary heterogeneity of charge to produce a structurally significant group of product ions.

MS Solutions #17: The Importance of Multiple-Charge Ion Precursors in Peptide MS/MS Sequencing-3
Figure 3: (a) Single-charge ion with proton sequestered on an R residue; (b) double-charge ion with one sequestered and one mobile proton.

A real-world example of this effect is shown in Figure 4. Electrospray spectra were collected for the synthetic peptide AFLDASR. Figure 4(a) shows the result of selection of the single-charge ion as the precursor. The lack of any significant number of structurally significant ions is a result of the sequestration of the charge on the C-terminal arginine. Figure 4(b) shows the same experiment except in this case the double-charge ion is selected as the precursor. We still see a y-ion dominated spectrum which we expect given the fixed charge on the C-terminus, but a reasonable number of ions are available for sequence determination. It is also important to note that the double-charge precursor ion is found in the middle of the mass-to-charge ratio range and that product ions are present which have both higher and lower m/z values than the precursor. This somewhat complicates interpretation, at least if done by hand calculation.

MS Solutions #17: The Importance of Multiple-Charge Ion Precursors in Peptide MS/MS Sequencing-4
Figure 4: Electrospray MS/MS experiments with the peptide AFLDASR: (a) single-charge precursor ion; (b) double-charge precursor ion.

Figure 5 shows a spectrum for the peptide AFLDASK, i.e., the only change in the sequence from the peptide in Figure 4 is substitution of a terminal lysine (K) in place of the arginine. The spectrum was obtained using a single-charge precursor and illustrates that lysine does not severely curtail proton mobility. A good variety of both N- and C-terminal ions are produced.
An interesting side-note to Figure 4 is production of a very abundant y 3 ion. This ion is produced by peptide bond cleavage to the C-terminal side of aspartic acid (D). This bond cleavage has been observed in a variety of aspartic acid-containing peptides with C-terminal arginines.

MS Solutions #17: The Importance of Multiple-Charge Ion Precursors in Peptide MS/MS Sequencing-5
Figure 5: Electrospray MS/MS experiment with the peptide AFLDASK using a single-charge precursor ion.

There are many sequence-dependent fragmentation patterns and more are being discovered all the time. The C-terminal arginine problem is fairly obvious but many, such as the aspartic acid example, are more subtle. It is therefore a good rule-of-thumb that if an inadequate number of structurally significant product ions are observed for a single-charge precursor then seek a multiple-charge precursor ion. While our focus here has been on peptide fragmentation, the same rule can be applied to other types of MS/MS experiments in which the analyte is capable of forming a multiple-charge ion.

Continue Learning With David Sparkman

Explore an in-depth applied mass spectrometry course from David Sparkman on Analytical Training Solutions. Strengthen fundamentals, improve spectral interpretation, and build confidence in day-to-day MS workflows across analytical applications.

View David Sparkman’s mass spectrometry course >>

Add Separation Science as a preferred source on Google

Add Separation Science as a preferred Google source to see more of our trusted coverage

Meet the Author(s):

  • David Sparkman

    O. David Sparkman is currently an Adjunct Professor of Chemistry at the University of the Pacific in Stockton, California; Contractor to the National Institute of Standards and Technology Mass Spectrometry Data Center; President of ChemUserWorld.com; and a former American Chemical Society Instructor (1978–2006) and American Society for Mass Spectrometry Member-at-large for Education (2004–2006). At the University of the Pacific, Prof. Sparkman teaches courses in mass spectrometry and analytical chemistry and manages the mass spectrometry facility. Over the past 28 years, he has developed and taught five different ACS courses in mass spectrometry; he holds positions on the Editorial Advisory Boards of the European Journal of Mass Spectrometry and the HD Science GC/MS Update – Part B; and is the Book Review Editor for the European Journal of Mass Spectrometry. He is the author of Mass Spectrometry Desk Reference (Global View Publishing: Pittsburgh, PA, 1st ed. 2000; 2nd ed. 2006). Prof. Sparkman is a member of the Editorial Boards of the John Wiley Encyclopedia of Environmental Analysis and Remediation and Encyclopedia of Analytical Chemistry, Editor of and a contributor to the Mass Spectrometry Section of the Encyclopedia of Analytical Chemistry, and a contributor to the Encyclopedia of Environmental Analysis and Remediation. Along with J. Throck Watson, he developed the Mass Spectral Interpretation Quick Reference Guide. He also provides general consulting service in mass spectrometry for a number of instrument manufacturers, manufacturing companies, and government agencies.

    View Full Profile