Glycans influence immune recognition, cellular communication, and interactions within the tumor microenvironment. Yet conventional analytical methods often lose the spatial information needed to understand their biological roles. Spatial glycomics addresses this limitation by mapping glycan distributions within intact tissue, revealing molecular patterns that other omics approaches may overlook. Separation Science spoke with Dr. Richard Drake, a researcher specializing in glycomics and mass spectrometry imaging at the Medical University of South Carolina, about the technical challenges of spatial glycan analysis, emerging applications in cancer research, and the barriers to clinical adoption.
Why is spatial glycomics important for understanding glycans in their native tumor environment?
Conventional liquid chromatography–mass spectrometry (LC-MS) provides detailed structural information, but tissue homogenization removes the spatial relationships between glycans and their surrounding environment. Matrix-assisted laser desorption/ionization (MALDI) mass spectrometry (MS) imaging allows us to examine these distributions within intact tissue sections. Early in our work, we found that glycan organization was far from random. Connective tissues and individual organs exhibited distinct glycosylation patterns. More recently, we have focused on immune microenvironments, including tertiary lymphoid structures, where glycan signatures may provide insight into local immune activity and responses to cancer immunotherapy. Spatial RNA transcriptomics cannot directly reveal post-translational modifications such as glycosylation. Glycans can also occur away from their cells of origin, particularly within the extracellular matrix. This makes spatial context important when interpreting their biological functions. Conventional imaging approaches have limitations as well. Antibodies are unavailable for many glycan structures, while lectins can recognize multiple carbohydrate motifs rather than individual glycan species. Glycan mass spectrometry imaging provides another way to investigate these distributions without relying exclusively on recognition reagents.
How do you prevent glycan delocalization during on-tissue MALDI imaging?
Formalin-fixed paraffin-embedded (FFPE) tissue offers an advantage because preparation aligns with established immunohistochemistry workflows, including dewaxing and antigen retrieval. One priority has been optimizing the enzymatic release of N-linked glycans. We use a highly active, stable peptide-N-glycosidase F (PNGase F PRIME from N-Zyme Scientifics) that operates without high-salt buffers, which can interfere with MALDI ionization and signal quality. Controlling humidity, sprayer airflow, and liquid application is equally important. Small core biopsy tissues, such as kidney or prostate samples, are particularly susceptible to delocalization because they lack the surface features of larger tissue sections that can help constrain diffusion. For these samples, we sometimes use automated matrix sublimation. This applies a matrix coating without introducing the air currents associated with conventional spraying, helping preserve spatial information. Quality control is also straightforward in principle: if PNGase F digestion fails, little or no released glycan signal appears. Under suitable acquisition conditions, we can assess digestion success and spatial resolution on small biopsies within approximately 30 minutes. Larger tissue sections may require 90 minutes to three hours. The challenge is balancing efficient glycan release, signal quality, and spatial resolution rather than optimizing any one parameter in isolation.
How does on-slide chemistry help distinguish structurally similar glycans?
Sialic acid linkages are particularly important because glycans with similar compositions can have different biological functions depending on how their sugars are connected. We adapted a double-amidation method originally developed by the Wuhrer lab in Leiden. The chemistry works with FFPE tissue and uses dimethyl sulfoxide (DMSO)-based conditions without requiring a specialized setup. The reactions distinguish alpha-2,3 and alpha-2,6 sialic acid linkages through characteristic mass shifts of approximately +37 Da and +27 Da, respectively. Our group has also modified the second amidation step to introduce chemical handles suitable for bioorthogonal click reactions. These modifications create opportunities to combine mass spectrometry imaging with complementary techniques, including avidin-based fluorescence microscopy on the same tissue section. We are interested in extending these approaches to other structures, including alpha-2,8-linked polysialic acids, which have relevance to neurological biology and certain cancers.
Can ion mobility resolve the remaining challenges of glycan isomer analysis?
Ion mobility provides another dimension of separation and can help distinguish structures that conventional high-resolution mass spectrometry cannot readily resolve. However, its usefulness depends partly on access to well-characterized isomeric standards, which unfortunately lacking for large N-linked glycan isomers. In the meantime, selective enzymatic treatments provide valuable structural information. For example, endoglycosidase F3 can help investigate core fucosylation. Under appropriate conditions, it produces a characteristic 349 Da mass difference that helps distinguish core from outer-arm fucosylation, although it does not resolve every possible configuration. We also use enzymes such as endo-beta-galactosidase, sometimes called LacNAcase, to break larger glycan structures into fragments that are easier to detect and characterize. These approaches complement ion mobility and high-resolution mass spectrometry. Looking ahead, spatial glycopeptidomics could provide another important advance by connecting individual glycan structures with their associated proteins within a tissue environment.
How can spatial glycomics contribute to multiomics research?
Our slide-based processing workflow can be adapted to different sample types, including serum, urine, and cerebrospinal fluid. Proteins can be immobilized on chemically functionalized slides, allowing us to remove salts and other interfering constituents before enzymatic treatment and MALDI analysis. In lupus nephritis research, we are exploring matched kidney biopsy, serum, and urine samples to compare glycan profiles across different biological compartments. This may help distinguish local tissue-associated changes from broader systemic disease processes. Combining glycomics with transcriptomics, proteomics, and metabolomics offers a more complete picture of disease biology. Each analytical approach provides different information, while glycomics adds insight into post-translational modifications and molecular organization that cannot be inferred directly from gene expression.
How could spatial glycomics advance immune-cell research?
Glycans are central to immune recognition, but their structural complexity and terminology has contributed to what some researchers call "glycophobia;" i.e., glycans are just too complex to study and understand. Immunologists often focus on transcripts, proteins, or cellular metabolism, even though immune cells interact extensively with carbohydrate structures through lectin receptors. Our group has developed a separate single-cell platform to investigate these interactions. Jake Dresman, a former student in our laboratory, developed an approach we call "flow on a stick." Rather than imaging individual immune cells within intact tissue, or use liquid flow cytometry separations, the method captures isolated cells on functionalized slides for molecular analysis. We can capture cells from peripheral blood mononuclear cell (PBMC) populations and analyze approximately six cells per second, allowing us to profile around 20,000 cells in under 30 minutes. The platform creates opportunities to examine glycans alongside lipids and metabolites across large immune-cell populations. One potential application is characterizing cellular therapeutics, including chimeric antigen receptor T-cell (CAR-T) products, where understanding molecular heterogeneity could provide valuable information. This remains an emerging application, but the ability to profile thousands of individual cells offers possibilities that would be difficult to achieve using lower-throughput methods.
What are the main barriers to bringing spatial glycomics into clinical pathology?
Instrumentation is not necessarily the biggest obstacle. MALDI mass spectrometry is already established for microbe identification in clinical microbiology labs, demonstrating that a MALDI assay can be incorporated into routine clinical workflows. Many sample-preparation procedures are manageable with appropriate training. In our laboratory, for example, summer students have learned established protocols within approximately two weeks. Clinical implementation, however, would require additional validation, quality control, and standardization. A greater challenge is interpretation. Glycobiology remains underrepresented in medical education, leaving many pathologists without the background needed to interpret complex glycan structures. Data management presents another obstacle. Spatial imaging can generate enormous sized datasets, particularly when multiple molecular measurements are combined or tissues are reconstructed in three dimensions. Artificial intelligence and computational tools may eventually help reduce these datasets to a smaller number of informative molecular features. Rather than requiring pathologists to interpret terabytes of imaging data, the goal would be to identify the patterns most relevant to disease. Broader adoption will depend on developing these interpretive capabilities alongside validated analytical workflows, potentially beginning in specialized clinical and research facilities.
Responses have been edited for length and clarity.