Column Equilibration in HILIC

by | May 14, 2019

The importance of HILIC column equilibration is discussed.

It is matter of fact that the samples most chromatographers have to face are rather complex for sample number and compound variety. In these situations, gradient elution is largely preferred for the separation of analytes with different retention factors. In this way it is possible to narrow the peak width of the later eluted compounds and shorten the run time. However, gradient separations have a principal drawback: after each run they require column equilibration with the starting concentration of the mobile phase before the next sample can be injected.

   In comparison to RPLC stationary phases, HILIC phases need gradients with lower slope and longer equilibration times. This is because in HILIC the separation is driven by the thickness of the water layer that originates from the water in the mobile phase. Changing the mobile phase composition will change the water layer. However, this is a slow process and requires several column volumes before the water in the water layer reaches equilibrium with the mobile phase. A HILIC mobile phase must always contain at least 2–3% of water in the final mobile phase in order to allow the formation of the water layer on the surface of stationary phase. Then a gradient elution should be performed increasing the water content up to a maximum of 40–50% in 20 minutes. Each HILIC column requires a different amount of column volume in order to reach a complete equilibration. Let’s consider a general column of dimensions 2.1 mm × 150 mm. The column volume is ~300 µL (for calculation of the column volume, see HPLC Solutions#101: Estimating Column Volume). HILIC columns need 20 column volumes, or more, for proper equilibration. This means that at a mobile phase flow-rate of 200–400 µL/min, the equilibration time will be between 30 and 15 minutes.

   Insufficient column equilibration will cause shifting of the retention times, resulting in poor reproducibility (Figure 1).

HILIC Solutions #8: Column Equilibration in HILIC
Figure 1: Retention time shift between runs due to insufficient column equilibration

   In order to shorten the equilibration time it is possible to perform the equilibration step at a flow-rate higher, even two‑fold, than the one used for the analysis. In HILIC this is possible owing to the low viscosity of the mobile phase at high organic content, which gives a low back pressure.

   Another possibility of increasing the analysis throughput shortening the equilibration step is to run the samples continually allowing a fast but constant equilibration time among the runs. In this case, the column reaches a dynamic equilibrium and even if there is not a complete equilibration, it is possible to obtain good reproducibility.

   In the next instalment, we will discuss the effect of sample diluent on HILIC retention and peak shape.

This blog article series is produced in collaboration with Dr Giorgia Greco, Product Manager with Thermo Fisher Scientific in Germany and Thomas Letzel, Associate Professor and Head of the Analytical Research Group at the Technische Universität München, Germany.
Giorgia Greco received a PhD in Chemistry and worked as a Post Doc researcher at the Technische Universität München, Germany. During her research, she specialized in the fundamental of LC-MS and in the separation and analysis of metabolites from human and food matrices, as well as organic contaminants in waste water samples, by hyphenated HPLC/MS and HILIC/MS techniques.
Thomas Letzel received his PhD in Chemistry with Aerosol Analysis and then worked as a Post-Doc performing pharmaceutical analysis. He is the author of more than 50 publications and two books and wants to share his experience in liquid chromatography, especially in HILIC, with the community to accelerate the dissemination about HILIC theory and practical handling.
If you have any questions about this article send them to techtips@sepscience.com

Related Content

Advertisment

Advertisment

Advertisement